Dental Turbine For Kavo 6000B

Dental Turbine For Kavo 6000B

Handpiece Turbine for Kavo 6000B

Also available:

  • Kavo 625CD
  • Kavo 634 F4 Companion
  • Kavo Super Torque 625/630/640
  • Kavo 647B/649B
  • Kavo 650B

Official PayPal Seal

Turbine Features

  • Economically priced to save you money over the life of your handpieces.
  • High speed precision balancing ensures silky smooth vibration-free operation.
  • Superior gripping force.
  • Easy push button operation.
  • Super Precision Dental Bearings (meet ISO Specifications) for very quiet operation.
  • Designed and manufactured for Superior-life, Superior torque, and to resist the negative effects of autoclaving.
  • Superior performance designed to be transparent, meet or exceeds OEM turbine.
  • Superior Spindles. Rebuildable spindles enable you to further save on maintenance costs.

Turbine Installation Guide

Step 1: Turbine Removal

Insert bur into the handpiece head. Unscrew back cap with appropriate back cap wrench. Remove turbine from the handpiece head by applying pressure to the bur.

Step 2: Inspect Handpiece

Examine handpiece head for damage or dents. If you observe damage, do not install the new turbine. Have the handpiece repaired by a qualified repair facility. The new turbine will be damaged if you proceed.

Step 3: Remove O-Rings and Other Components

Remove any remaining o'rings out of the handpiece back cap and inside the handpiece head. Remove any other existing parts as well, such as wave washers, shims, spacers, or broken bearing parts. Use an explorer to insure parts are removed, as they often hide in blackened debris.

Step 4: Clean Handpiece

If possible, ultra sonic clean the handpiece shell with a handpiece friendly degreaser for 15 minutes, flush handpiece when done. If you do not use an ultrasonic cleaner, spray handpiece cleaner into the handpiece head. Use a cotton swab to remove any debris left in the head. Using air, blow out the handpiece head to remove any cotton fibers.

Step 5: Attach Back Cap

As a general rule, you should attach the back cap onto the new turbine (if not already attached). If it does not attach easily, place a drop of oil and spread it around the back o'ring/bearing. Carefully insert the back bearing into the back cap, gently rocking it from side to side. Too much force will cut the o'ring or damage the back bearing.

Step 6: Install Preparation

Make sure that new o'rings, wave washers, and spacers are seated in the proper position on the turbine or in the handpiece head. These parts and their positioning depend on your particular handpiece/turbine make and model. For example, if you had to remove an o'ring out of a groove in the handpiece head in Step 3, then you will need to put a new o'ring in its place at this time.

Step 7: Turbine Installation

Holding the turbine upright (bur pointed up), insert turbine into the handpiece head. Thread back cap into the handpiece until finger tight. Grasp the shipping bur and gently move the turbine in a side to side motion to help seat the o'ring. If any resistance occurs while threading the back cap, unscrew 1/4 turn and then continue to thread. Rotate shipping bur to ensure that the turbine moves freely. If not, unscrew back cap and repeat this step. Never force the back cap if resistance occurs.

Step 8: Lubricate Turbine (For Lube Type Turbines Only)

Place 3 drops of oil in the drive air tube and operate handpiece.

Troubleshooting after Turbine Installation

1. The turbine does not spin freely after installation

This is usually caused by improper installation. Most likely, the turbine has rolled, moved or damaged the front o-ring during installation. Sometimes, this occurs when old components are left in the handpiece head during turbine removal. In cases where springs and/or shims are placed on front of the turbine during installation, they could settle incorrectly and cause resistance. In all cases, re-install the turbine, do not run the handpiece until corrected.

2. The turbine sounds loud or runs rough after installation

This is usually caused by o-ring misplacement or damage, or improper back cap placement. In these cases, re-install the turbine. It is very important to replace o-rings with new ones, worn o-rings could cause this problem. Also, debris left in the handpiece head or head damage could cause this problem. Try removing the turbine and cleaning the handpiece as described in the installation instructions. If the handpiece head has damage or dents, the handpiece must be repaired before using. Damaged handpiece heads will damage turbines.

3. The back cap gets hot during use

In these cases, stop using the handpiece. Re-install the turbine to ensure that it is seated properly. If the turbine is not assembled in the proper position on the spindle, or has been moved, the back of the spindle could rub against the back cap. Continued use will damage the turbine and could cause discomfort to patients. Also, your push button back cap could be stuck in the down position, which would then rub against the spindle. In this case, remove the back cap and clean. If the back cap continues to stick, it must be replaced.

4. I'm having turbine failures

Improper maintenance procedures is the number one reason for turbine failures. Do not over temperature your handpieces during sterilization cycles (not to exceed 275 f). Handpieces should be flushed prior to sterilization to remove debris. Handpieces should be lubricated after sterilization (maintenance free turbines should be lubricated from time to time to lubricate components such as o-rings and spindle assemblies).

5. My handpieces are running slow

Try removing the turbine and cleaning the inside of the handpiece head, debris could cause slow performance. Check to ensure that your air pressure is properly gauged at the chair unit. Try replacing the gasket at the back end of the handpiece, a worn gasket will cause air to escape out the back of the handpiece.

6. Warranty

If your turbine fails within the warranty period due to defects in materials or workmanship you may be entitled to a replacement unit.

We offer the most complete line of dental handpiece parts and products available. As you will clearly see, we provide our customers the most competitive pricing in the dental industry. Our products will meet or exceed ISO specifications.

We help our customers (dentists, handpiece repair personals) improve their business by offering the highest quality products at the reasonable prices. All items our expertly assemble and must pass through quality control procedures before being shipped. This insures that the you received a superior quality product each and every time an order is placed.

An after-market part or product is a new part or product that is manufactured by someone other than the original equipment manufacture that either meets or exceeds the same stringent guidelines and or specifications of the OEM.

The benefit of using an after-market product is that you are able to get the same quality product at a substantially lower cost than if you had purchased an OEM product. With the savings in bottom line costs, this allows a higher return on investment without having to compromise durability and product quality.

"The sale of this item may be subject to strict regulation by the U.S. FDA and state and local regulatory agencies. If so do not bid on this item unless you are an authorized purchaser. If the item is subject to FDA regulations I will verify your status as an authorized purchaser of this item before shipment of the item."

Disclaimer

Crown Trading Inc. provides high quality replacement turbines and turbine parts from multiple dental product manufacturers. Our competitive warranties equal industry standards.

Crown Trading Inc. makes no insuation that any of its products it sells or uses in repairs are from the original handpiece factory for the below listed registered trade names or that any of the companies listed below endorse our products:

Midwest; XGT, Quiet Air, Stylus, Tradition, 8000, Shorty, Star; Vista, 430, Advantage, Solara, Concentric, Futura, Ultra, Titan, Kavo; Contact Air, Bella Torque, Super Torque, Mira Lux, Lares; 557, 757, NSK; Mach Lite, Phatelus, Viper, Push & Swivel, Super Grade, Pana Air, W&H / A-Dec Trend, Synea, Bien Air Ondine, Black Pearl, Prestige, Bora, Lab, Impact Air 45, Siemens, Sirona, Dentex, Kinetic, Viper, Quantum, Dabi Atlante, Silent, Micro Motors, MTI Lynx, Champion, Athena, Healthco, Johnson-Promident, Naminex, Pana-Air, Patterson, Safco, Schein, Super-Trac, True Speed, Score, EZ Press III

Price:  $139.95